Comparison of central corneal thickness measurements by Galilei Dual-Scheimpflug Analyzer® and ultrasound pachymeter in myopic eyes

ArticleAçık Erişim

Özet

■ BACKGROUND AND OBJECTIVE: To compare the central corneal thickness (CCT) measurements by Galilei Dual-Scheimpflug analyzer (GSA) (Ziemer Group, Port, Switzerland) and ultrasound pachymeter (UP) in myopic eyes. ■ PATIENTS AND METHODS: In this prospective study, CCT of 161 myopic eyes of 81 refractive surgery candidates (35 female, 46 male; mean age: 23.18 ± 4.08 years) were measured by GSA and UP consecutively. The data were analyzed statistically by paired t test, Bland-Altman plot, and Pearson correlation test to assess the agreement of the measurements. ■ RESULTS: The mean CCTs obtained by GSA and UP were 559.85 ± 30.87 and 560.41 ± 34.45 μm, respectively. No significant difference was found between the measurements (P = .684) and regression analysis showed a high correlation between the measurements obtained with both devices (r = 0.86; P < .001). A strong agreement between the two methods was also found by Bland-Altman plot. ■ CONCLUSION: In myopic eyes, the CCT measurements of GSA and UP are similar and highly correlated. Because of high agreement between these devices, the GSA is a non-contact method that may be an alternative to UP for measurement of CCT. Copyright © SLACK Incorporated.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/15518Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorYeter, Volkan
dc.contributor.authorSönmez, Baris
dc.contributor.authorBeden, Ümit
dc.date.accessioned2026-10-18T20:30:21.743900806Z
dc.date.available2026-10-18T20:30:21.743900806Z
dc.date.issued2012
dc.description.abstract■ BACKGROUND AND OBJECTIVE: To compare the central corneal thickness (CCT) measurements by Galilei Dual-Scheimpflug analyzer (GSA) (Ziemer Group, Port, Switzerland) and ultrasound pachymeter (UP) in myopic eyes. ■ PATIENTS AND METHODS: In this prospective study, CCT of 161 myopic eyes of 81 refractive surgery candidates (35 female, 46 male; mean age: 23.18 ± 4.08 years) were measured by GSA and UP consecutively. The data were analyzed statistically by paired t test, Bland-Altman plot, and Pearson correlation test to assess the agreement of the measurements. ■ RESULTS: The mean CCTs obtained by GSA and UP were 559.85 ± 30.87 and 560.41 ± 34.45 μm, respectively. No significant difference was found between the measurements (P = .684) and regression analysis showed a high correlation between the measurements obtained with both devices (r = 0.86; P < .001). A strong agreement between the two methods was also found by Bland-Altman plot. ■ CONCLUSION: In myopic eyes, the CCT measurements of GSA and UP are similar and highly correlated. Because of high agreement between these devices, the GSA is a non-contact method that may be an alternative to UP for measurement of CCT. Copyright © SLACK Incorporated.
dc.identifier.doi10.3928/15428877-20120102-08
dc.identifier.endpage134
dc.identifier.issn1542-8877
dc.identifier.issue2
dc.identifier.pmid22320416
dc.identifier.scopus2-s2.0-84859811227
dc.identifier.startpage128
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/15518
dc.identifier.volume43
dc.language.isoen
dc.relation.ispartofOphthalmic Surgery Lasers and Imaging
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectAdolescent
dc.subjectAdult
dc.subjectCornea
dc.subjectDiagnostic Techniques, Ophthalmological
dc.subjectFemale
dc.subjectHumans
dc.subjectMale
dc.subjectMyopia
dc.subjectPhotography
dc.subjectProspective Studies
dc.subjectRegression Analysis
dc.subjectUltrasonography
dc.subjectYoung Adult
dc.subjectanalytical equipment
dc.subjectanalyzer
dc.subjectarticle
dc.subjectbland altman plot
dc.subjectcontrolled study
dc.subjectcornea thickness
dc.subjectcorrelation coefficient
dc.subjectdata analysis
dc.subjecthuman
dc.subjectmajor clinical study
dc.subjectpriority journal
dc.subjectprospective study
dc.subjectrefractive surgery
dc.subjectstatistical analysis
dc.subjectStudent t test
dc.subjectultrasound pachymeter
dc.titleComparison of central corneal thickness measurements by Galilei Dual-Scheimpflug Analyzer® and ultrasound pachymeter in myopic eyes
dc.typeArticle
dspace.entity.typePublication
rdl.citation.scopus13
rdl.departmentDepartment of Ophthalmology, Erzincan University Mengücek Gazi Education and Research Hospital, Erzincan, Turkey. dr.volkanyeter@gmail.com
rdl.source.indexSCOPUS
rdl.source.indexPUBMED