New PVC Membrane-Based Ion-Selective Electrode for Ultrasensitive Detection of Bismuth(III) with Sudan I

ArticleAçık Erişim

Özet

Accurate and dependable measurement of bismuth is vital for ensuring the safety of drugs, managing industrial processes, and evaluating environmental pollution. In this study, a new potentiometric sensor that uses Sudan I as the electroactive material within poly(vinyl chloride) (PVC) membrane structure was designed. The sensor displayed a broad linear concentration range of 1.0 ×10-3 to 1.0 ×10-10 M, an ultra-low detection limit of 4.0 ×10-11 M, and a response time of 30 seconds. It retained stable performance across a pH range of 2.0–7.0 and had an operational lifetime of about one month. When compared to existing literature, the sensor's selectivity and detection limit were significantly enhanced. It was successfully utilized to determine bismuth species in pharmaceutical formulations. Potentiometric methods provide clear advantages over other analytical techniques, such as simplicity, cost-effectiveness, quick response, and high selectivity. This developed sensor is a valuable tool for bismuth(III) determination, especially in quality control and pharmaceutical analysis. © 2025, Chemical Society of Pakistan. All rights resevered.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/10683Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorAksar, Burak
dc.contributor.authorKanberoglu, Gulsh Saydan
dc.contributor.authorColdur, Fatih
dc.date.accessioned2026-10-18T20:25:51.046851241Z
dc.date.available2026-10-18T20:25:51.046851241Z
dc.date.issued2025
dc.description.abstractAccurate and dependable measurement of bismuth is vital for ensuring the safety of drugs, managing industrial processes, and evaluating environmental pollution. In this study, a new potentiometric sensor that uses Sudan I as the electroactive material within poly(vinyl chloride) (PVC) membrane structure was designed. The sensor displayed a broad linear concentration range of 1.0 ×10-3 to 1.0 ×10-10 M, an ultra-low detection limit of 4.0 ×10-11 M, and a response time of 30 seconds. It retained stable performance across a pH range of 2.0–7.0 and had an operational lifetime of about one month. When compared to existing literature, the sensor's selectivity and detection limit were significantly enhanced. It was successfully utilized to determine bismuth species in pharmaceutical formulations. Potentiometric methods provide clear advantages over other analytical techniques, such as simplicity, cost-effectiveness, quick response, and high selectivity. This developed sensor is a valuable tool for bismuth(III) determination, especially in quality control and pharmaceutical analysis. © 2025, Chemical Society of Pakistan. All rights resevered.
dc.description.sponsorshipThe present study was funded by Van Yuzuncu Yil University Scientific Research Projects Coordination Unit under Grant number FYL-2024-11280.
dc.identifier.doi10.52568/001798/jcsp/47.05.2025
dc.identifier.issn0253-5106
dc.identifier.issue5
dc.identifier.scopus2-s2.0-105031940501
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/10683
dc.identifier.volume47
dc.language.isoen
dc.publisherChemical Society of Pakistan
dc.relation.ispartofJournal of the Chemical Society of Pakistan
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectBi(III)-selective sensor
dc.subjectIon-selective electrode
dc.subjectPharmaceutical analysis
dc.subjectPotentiometric determination
dc.subjectPVC membrane electrode
dc.titleNew PVC Membrane-Based Ion-Selective Electrode for Ultrasensitive Detection of Bismuth(III) with Sudan I
dc.typeArticle
dspace.entity.typePublication
rdl.citation.scopus0
rdl.oa.statusopen
rdl.source.indexSCOPUS