Development of a PVC Membrane Potentiometric Sensor with Low Detection Limit and Wide Linear Range for the Determination of Maprotiline in Pharmaceutical Formulations
Özet
Maprotiline (MAP), is an active ingredient of some antidepressants which has been broadly used for the therapy of depression. The overdose of MAP in the treatment of patients with depression can lead to significant side effects, even lethal intoxication. Therefore, it is very important to develop simple, inexpensive, fast and reliable detection methods for MAP in interested samples. In our present study, we have offered a novel poly(vinylchloride) (PVC) membrane maprotiline-selective potentiometric sensor based on a type of MOFs, known as MIL-53(Al), as electroactive ionophore substance. The sensor prepared at the optimum membrane composition exhibited a linear response for maprotiline hydrochloride in the concentration range of 1,0×10−9 M−1,0×10−2 M with a slope of 59,1 mV/decade and detection limit of 9,0×10−10 M. The pH working range of the sensor was determined as 2,0–9,5. The response time of the sensor was very fast and less than 5 s. With these performance features, it has been shown that the current sensor has quite superior features compared to the potentiometric MAP-selective sensor available in the literature. The potentiometric application of the sensor was successfully carried out with the detection of MAP in pharmaceutical tablets containing MAP-HCl. © 2022 Wiley-VCH GmbH
Anahtar Kelimeler
Tam Kayıt
| Alan | Değer | Dil |
|---|---|---|
| dc.contributor.author | Tekce, Serkan | |
| dc.contributor.author | Subasi, Yaver | |
| dc.contributor.author | Coldur, Fatih | |
| dc.contributor.author | Kanberoglu, Gulsah Saydan | |
| dc.contributor.author | Zahmakiran, Mehmet | |
| dc.date.accessioned | 2026-10-18T20:27:32.281469489Z | |
| dc.date.available | 2026-10-18T20:27:32.281469489Z | |
| dc.date.issued | 2022 | |
| dc.description.abstract | Maprotiline (MAP), is an active ingredient of some antidepressants which has been broadly used for the therapy of depression. The overdose of MAP in the treatment of patients with depression can lead to significant side effects, even lethal intoxication. Therefore, it is very important to develop simple, inexpensive, fast and reliable detection methods for MAP in interested samples. In our present study, we have offered a novel poly(vinylchloride) (PVC) membrane maprotiline-selective potentiometric sensor based on a type of MOFs, known as MIL-53(Al), as electroactive ionophore substance. The sensor prepared at the optimum membrane composition exhibited a linear response for maprotiline hydrochloride in the concentration range of 1,0×10−9 M−1,0×10−2 M with a slope of 59,1 mV/decade and detection limit of 9,0×10−10 M. The pH working range of the sensor was determined as 2,0–9,5. The response time of the sensor was very fast and less than 5 s. With these performance features, it has been shown that the current sensor has quite superior features compared to the potentiometric MAP-selective sensor available in the literature. The potentiometric application of the sensor was successfully carried out with the detection of MAP in pharmaceutical tablets containing MAP-HCl. © 2022 Wiley-VCH GmbH | |
| dc.description.sponsorship | We express our deep gratitude to Faculty of Science Deanery (Van Yuzuncu Yil University) and Faculty of Arts and Sciences Deanery (Erzincan Binali Yildirim University) for the facilities and opportunities they provided us to conduct the current study. | |
| dc.identifier.doi | 10.1002/slct.202103553 | |
| dc.identifier.issn | 2365-6549 | |
| dc.identifier.issue | 2 | |
| dc.identifier.scopus | 2-s2.0-85123692359 | |
| dc.identifier.startpage | e202103553 | |
| dc.identifier.uri | https://rdlopen.rawdatalibrary.net/r/ebyu/12660 | |
| dc.identifier.volume | 7 | |
| dc.identifier.wos | WOS:000753970000013 | |
| dc.language.iso | en | |
| dc.publisher | John Wiley and Sons Inc | |
| dc.relation.ispartof | ChemistrySelect | |
| dc.rights | info:eu-repo/semantics/openAccess | |
| dc.subject | maprotiline | |
| dc.subject | metal-organic frameworks | |
| dc.subject | pharmaceutical analysis | |
| dc.subject | pvc membrane electrode | |
| dc.subject | sensors | |
| dc.subject | PERFORMANCE LIQUID-CHROMATOGRAPHY | |
| dc.subject | ION-SELECTIVE ELECTRODES | |
| dc.subject | SOLID-PHASE EXTRACTION | |
| dc.subject | RECENT PROGRESS | |
| dc.subject | F-MOFS | |
| dc.subject | WATER | |
| dc.subject | ADSORPTION | |
| dc.subject | REMOVAL | |
| dc.subject | DESIPRAMINE | |
| dc.subject | Chemistry, Multidisciplinary | |
| dc.subject | Chemistry | |
| dc.title | Development of a PVC Membrane Potentiometric Sensor with Low Detection Limit and Wide Linear Range for the Determination of Maprotiline in Pharmaceutical Formulations | |
| dc.type | Article | |
| dspace.entity.type | Publication | |
| rdl.citation.scopus | 7 | |
| rdl.citation.wos | 7 | |
| rdl.source.index | SCOPUS | |
| rdl.source.index | WOS | |
| rdl.wos.index | Science Citation Index Expanded (SCI-EXPANDED) |