A novel borate ion selective electrode based on carbon nanotube-silver borate

ArticleAçık Erişim

Özet

A novel miniaturized borate selective composite membrane electrode based on carbon nanotube-silver borate is described. The developed electrode exhibited a stable and linear potentiometric behavior over the concentration range of 1 x 10-1 M and 1 x 10-4 M borate ion with a Nernstian slope of 34 mV/decade and a limit of detection of 2 x 10-5 M. The electrode showed a clear selectivity to borate ion over sulfate, chloride, nitrate and many other inorganic acid anions. The response time of the electrode was found less than 14 s. Potentiometric readings of the electrode were consistent, when pH was between 4 and 8, through almost 3 months. The developed borate-selective composite membrane electrode was directly applied for the quantification of borate ions in real water samples. The results obtained by the developed electrode method was statistically analyzed and compared with those obtained by ICP-MS method. No significant differences for either accuracy or precision were observed. © 2020 The Authors.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/13966Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.20964/2020.01.40Yayıncı resmi DOI adresi
Tam Kayıt
AlanDeğerDil
dc.contributor.authorBahar, Duygu Uner
dc.contributor.authorTopcu, Cihan
dc.contributor.authorOzcimen, Didem
dc.contributor.authorIsildak, Ibrahim
dc.date.accessioned2026-10-18T20:28:46.009809413Z
dc.date.available2026-10-18T20:28:46.009809413Z
dc.date.issued2020
dc.description.abstractA novel miniaturized borate selective composite membrane electrode based on carbon nanotube-silver borate is described. The developed electrode exhibited a stable and linear potentiometric behavior over the concentration range of 1 x 10-1 M and 1 x 10-4 M borate ion with a Nernstian slope of 34 mV/decade and a limit of detection of 2 x 10-5 M. The electrode showed a clear selectivity to borate ion over sulfate, chloride, nitrate and many other inorganic acid anions. The response time of the electrode was found less than 14 s. Potentiometric readings of the electrode were consistent, when pH was between 4 and 8, through almost 3 months. The developed borate-selective composite membrane electrode was directly applied for the quantification of borate ions in real water samples. The results obtained by the developed electrode method was statistically analyzed and compared with those obtained by ICP-MS method. No significant differences for either accuracy or precision were observed. © 2020 The Authors.
dc.identifier.doi10.20964/2020.01.40
dc.identifier.endpage914
dc.identifier.issn1452-3981
dc.identifier.issue1
dc.identifier.scopus2-s2.0-85082137653
dc.identifier.startpage899
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/13966
dc.identifier.volume15
dc.identifier.wosWOS:000505155000075
dc.language.isoen
dc.publisherElectrochemical Science Group
dc.relation.ispartofInternational Journal of Electrochemical Science
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectBorate in environmental waters
dc.subjectBorate selective composite membrane electrode
dc.subjectCarbon nanotube-silver borate
dc.subjectPotentiometry
dc.titleA novel borate ion selective electrode based on carbon nanotube-silver borate