Comparison of central corneal thickness measurements by Galilei Dual-Scheimpflug Analyzer® and ultrasound pachymeter in myopic eyes
Özet
■ BACKGROUND AND OBJECTIVE: To compare the central corneal thickness (CCT) measurements by Galilei Dual-Scheimpflug analyzer (GSA) (Ziemer Group, Port, Switzerland) and ultrasound pachymeter (UP) in myopic eyes. ■ PATIENTS AND METHODS: In this prospective study, CCT of 161 myopic eyes of 81 refractive surgery candidates (35 female, 46 male; mean age: 23.18 ± 4.08 years) were measured by GSA and UP consecutively. The data were analyzed statistically by paired t test, Bland-Altman plot, and Pearson correlation test to assess the agreement of the measurements. ■ RESULTS: The mean CCTs obtained by GSA and UP were 559.85 ± 30.87 and 560.41 ± 34.45 μm, respectively. No significant difference was found between the measurements (P = .684) and regression analysis showed a high correlation between the measurements obtained with both devices (r = 0.86; P < .001). A strong agreement between the two methods was also found by Bland-Altman plot. ■ CONCLUSION: In myopic eyes, the CCT measurements of GSA and UP are similar and highly correlated. Because of high agreement between these devices, the GSA is a non-contact method that may be an alternative to UP for measurement of CCT. Copyright © SLACK Incorporated.
Anahtar Kelimeler
Tam Kayıt
| Alan | Değer | Dil |
|---|---|---|
| dc.contributor.author | Yeter, Volkan | |
| dc.contributor.author | Sönmez, Baris | |
| dc.contributor.author | Beden, Ümit | |
| dc.date.accessioned | 2026-10-18T20:30:21.743900806Z | |
| dc.date.available | 2026-10-18T20:30:21.743900806Z | |
| dc.date.issued | 2012 | |
| dc.description.abstract | ■ BACKGROUND AND OBJECTIVE: To compare the central corneal thickness (CCT) measurements by Galilei Dual-Scheimpflug analyzer (GSA) (Ziemer Group, Port, Switzerland) and ultrasound pachymeter (UP) in myopic eyes. ■ PATIENTS AND METHODS: In this prospective study, CCT of 161 myopic eyes of 81 refractive surgery candidates (35 female, 46 male; mean age: 23.18 ± 4.08 years) were measured by GSA and UP consecutively. The data were analyzed statistically by paired t test, Bland-Altman plot, and Pearson correlation test to assess the agreement of the measurements. ■ RESULTS: The mean CCTs obtained by GSA and UP were 559.85 ± 30.87 and 560.41 ± 34.45 μm, respectively. No significant difference was found between the measurements (P = .684) and regression analysis showed a high correlation between the measurements obtained with both devices (r = 0.86; P < .001). A strong agreement between the two methods was also found by Bland-Altman plot. ■ CONCLUSION: In myopic eyes, the CCT measurements of GSA and UP are similar and highly correlated. Because of high agreement between these devices, the GSA is a non-contact method that may be an alternative to UP for measurement of CCT. Copyright © SLACK Incorporated. | |
| dc.identifier.doi | 10.3928/15428877-20120102-08 | |
| dc.identifier.endpage | 134 | |
| dc.identifier.issn | 1542-8877 | |
| dc.identifier.issue | 2 | |
| dc.identifier.pmid | 22320416 | |
| dc.identifier.scopus | 2-s2.0-84859811227 | |
| dc.identifier.startpage | 128 | |
| dc.identifier.uri | https://rdlopen.rawdatalibrary.net/r/ebyu/15518 | |
| dc.identifier.volume | 43 | |
| dc.language.iso | en | |
| dc.relation.ispartof | Ophthalmic Surgery Lasers and Imaging | |
| dc.rights | info:eu-repo/semantics/openAccess | |
| dc.subject | Adolescent | |
| dc.subject | Adult | |
| dc.subject | Cornea | |
| dc.subject | Diagnostic Techniques, Ophthalmological | |
| dc.subject | Female | |
| dc.subject | Humans | |
| dc.subject | Male | |
| dc.subject | Myopia | |
| dc.subject | Photography | |
| dc.subject | Prospective Studies | |
| dc.subject | Regression Analysis | |
| dc.subject | Ultrasonography | |
| dc.subject | Young Adult | |
| dc.subject | analytical equipment | |
| dc.subject | analyzer | |
| dc.subject | article | |
| dc.subject | bland altman plot | |
| dc.subject | controlled study | |
| dc.subject | cornea thickness | |
| dc.subject | correlation coefficient | |
| dc.subject | data analysis | |
| dc.subject | human | |
| dc.subject | major clinical study | |
| dc.subject | priority journal | |
| dc.subject | prospective study | |
| dc.subject | refractive surgery | |
| dc.subject | statistical analysis | |
| dc.subject | Student t test | |
| dc.subject | ultrasound pachymeter | |
| dc.title | Comparison of central corneal thickness measurements by Galilei Dual-Scheimpflug Analyzer® and ultrasound pachymeter in myopic eyes | |
| dc.type | Article | |
| dspace.entity.type | Publication | |
| rdl.citation.scopus | 13 | |
| rdl.department | Department of Ophthalmology, Erzincan University Mengücek Gazi Education and Research Hospital, Erzincan, Turkey. dr.volkanyeter@gmail.com | |
| rdl.source.index | SCOPUS | |
| rdl.source.index | PUBMED |