Comparison of normal and keratoconic corneas by Galilei Dual-Scheimpflug Analyzer

ArticleAçık Erişim

Özet

Background and objective: To determine the efficacy of different Galilei Scheimpflug-Analyzer (GSA) parameters in discriminating between keratoconic and myopic eyes. Patients and methods: GSA measurements were obtained for 67 patients (67 eyes) with keratoconus and 151 patients (151 eyes) with myopia or myopic astigmatism. Several parameters, provided by the software or derived from the elevation maps, were evaluated and compared for the two groups. Results: Between the two groups, statistically significant differences were observed for all corneal parameters obtained by GSA (P < 0.001) except for the anterior chamber depth (P= 0.149). ROC analysis determined that posterior corneal elevation was the best predictive parameter (area under the curve: 0.99). The posterior corneal elevation, at a cut-off value of 18.5. μm, had 98.5% sensitivity and 98.3% specificity in discriminating keratoconus from myopic eyes. Conclusion: Elevation, pachymetric and keratometric parameters measured by the GSA, as well as the specific predictive GSA software parameters can effectively distinguish advanced keratoconus from myopic corneas. Also, keratoconus that is easily diagnosed by other means can be diagnosed easily by GSA software parameters. © 2013 British Contact Lens Association.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/15862Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.1016/j.clae.2013.04.001Yayıncı resmi DOI adresi
Tam Kayıt
AlanDeğerDil
dc.contributor.authorDemir, Selim
dc.contributor.authorSönmez, Bariş
dc.contributor.authorYeter, Volkan
dc.contributor.authorOrtak, Hüseyin
dc.date.accessioned2026-10-18T20:30:42.624231229Z
dc.date.available2026-10-18T20:30:42.624231229Z
dc.date.issued2013
dc.description.abstractBackground and objective: To determine the efficacy of different Galilei Scheimpflug-Analyzer (GSA) parameters in discriminating between keratoconic and myopic eyes. Patients and methods: GSA measurements were obtained for 67 patients (67 eyes) with keratoconus and 151 patients (151 eyes) with myopia or myopic astigmatism. Several parameters, provided by the software or derived from the elevation maps, were evaluated and compared for the two groups. Results: Between the two groups, statistically significant differences were observed for all corneal parameters obtained by GSA (P < 0.001) except for the anterior chamber depth (P= 0.149). ROC analysis determined that posterior corneal elevation was the best predictive parameter (area under the curve: 0.99). The posterior corneal elevation, at a cut-off value of 18.5. μm, had 98.5% sensitivity and 98.3% specificity in discriminating keratoconus from myopic eyes. Conclusion: Elevation, pachymetric and keratometric parameters measured by the GSA, as well as the specific predictive GSA software parameters can effectively distinguish advanced keratoconus from myopic corneas. Also, keratoconus that is easily diagnosed by other means can be diagnosed easily by GSA software parameters. © 2013 British Contact Lens Association.
dc.identifier.doi10.1016/j.clae.2013.04.001
dc.identifier.endpage225
dc.identifier.issn1476-5411
dc.identifier.issue5
dc.identifier.pmid23642799
dc.identifier.scopus2-s2.0-84882274804
dc.identifier.startpage219
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/15862
dc.identifier.volume36
dc.identifier.wosWOS:000324450000003
dc.language.isoen
dc.publisherELSEVIER SCIENCE BV
dc.relation.ispartofContact Lens and Anterior Eye
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectAnterior segment imaging
dc.subjectClinical examination
dc.subjectGalilei
dc.subjectKeratoconus
dc.subjectScheimpflug
dc.subjectAdult
dc.subjectCornea
dc.subjectCorneal Topography
dc.subjectDiagnosis, Differential
dc.subjectEquipment Design
dc.subjectEquipment Failure Analysis
dc.subjectFemale
dc.subjectHumans
dc.subjectMale
dc.subjectMyopia
dc.subjectPhotography
dc.subjectReproducibility of Results
dc.subjectSensitivity and Specificity
dc.subjectanalyzer
dc.subjectanterior eye chamber depth
dc.subjectarea under the curve
dc.subjectarticle
dc.subjectastigmatism
dc.subjectclinical effectiveness
dc.subjectcomputer program
dc.subjectcontrolled study
dc.subjectdiagnostic test accuracy study
dc.subjecthuman
dc.subjectmajor clinical study
dc.subjectpriority journal
dc.titleComparison of normal and keratoconic corneas by Galilei Dual-Scheimpflug Analyzer
dc.typeArticle
dspace.entity.typePublication
rdl.citation.scopus16
rdl.citation.wos10
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)