Solid-contact polyvinyl chloride membrane electrode based on the bis[(2-(hydroxyethylimino)phenolato]copper(II) complex for trace level determination of copper ions in wastewater

ArticleAçık Erişim

Özet

A miniaturized solid-contact polyvinyl chloride membrane electrode based on the bis[(2-(hydroxyethylimino)phenolato]copper(II) complex has been prepared for the determination of Cu(II) ion in real samples. The influence of coating membrane composition, pH, and possible interfering cations on the response of the ion-selective electrode was investigated. A number of membranes at different compositions were prepared and their Cu2+ sensing capabilities were investigated. The best results for Cu2+ sensing was obtained for the electrode membrane containing ionophore - 2-nitrophenyloctyl ether-polyvinyl chloride and potassium tetrakis(p-chlorophenyl)borate in a ratio of 4:65:30:1 (w/w, mg), respectively. The prepared electrode exhibited a near-Nernstian response (28.3 mV/decade) to Cu2+ in the concentration range from 1.0 × 10-6 to 1.0 × 10-1 mol/L with a limit of detection of 8.3 × 10-7 mol/L. The potentiometric response of the electrode was independent from the pH of the test solution in the pH range of 3.0-6.0. The electrode has a very short response time, about 15 s, and it can be used over a period of 3 months without any divergence in potential. The electrode has been successfully applied as an indicator electrode in potentiometric titration of Cu(II) ions with ethylenediaminetetraacetic acid and in the determination of Cu(II) in wastewater of the copper processing industry. © 2014 Published by NRC Research Press.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/15612Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.1139/cjc-2013-0530Yayıncı resmi DOI adresi
Tam Kayıt
AlanDeğerDil
dc.contributor.authorAndac, Muberra
dc.contributor.authorColdur, Fatih
dc.contributor.authorBilir, Seda
dc.contributor.authorBirinci, Aysenur
dc.contributor.authorDemir, Serkan
dc.contributor.authorUzun, Harun
dc.date.accessioned2026-10-18T20:30:27.692133146Z
dc.date.available2026-10-18T20:30:27.692133146Z
dc.date.issued2014
dc.description.abstractA miniaturized solid-contact polyvinyl chloride membrane electrode based on the bis[(2-(hydroxyethylimino)phenolato]copper(II) complex has been prepared for the determination of Cu(II) ion in real samples. The influence of coating membrane composition, pH, and possible interfering cations on the response of the ion-selective electrode was investigated. A number of membranes at different compositions were prepared and their Cu2+ sensing capabilities were investigated. The best results for Cu2+ sensing was obtained for the electrode membrane containing ionophore - 2-nitrophenyloctyl ether-polyvinyl chloride and potassium tetrakis(p-chlorophenyl)borate in a ratio of 4:65:30:1 (w/w, mg), respectively. The prepared electrode exhibited a near-Nernstian response (28.3 mV/decade) to Cu2+ in the concentration range from 1.0 × 10-6 to 1.0 × 10-1 mol/L with a limit of detection of 8.3 × 10-7 mol/L. The potentiometric response of the electrode was independent from the pH of the test solution in the pH range of 3.0-6.0. The electrode has a very short response time, about 15 s, and it can be used over a period of 3 months without any divergence in potential. The electrode has been successfully applied as an indicator electrode in potentiometric titration of Cu(II) ions with ethylenediaminetetraacetic acid and in the determination of Cu(II) in wastewater of the copper processing industry. © 2014 Published by NRC Research Press.
dc.description.sponsorshipThis work was supported by the Ondokuz Mayis University Research Foundation (BAP) Project (PYO.Fen.1904.10.034).
dc.identifier.doi10.1139/cjc-2013-0530
dc.identifier.eissn1480-3291
dc.identifier.endpage328
dc.identifier.issn0008-4042
dc.identifier.issue4
dc.identifier.scopus2-s2.0-84898077817
dc.identifier.startpage324
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/15612
dc.identifier.volume92
dc.identifier.wosWOS:000334031500009
dc.language.isoen
dc.publisherCANADIAN SCIENCE PUBLISHING
dc.relation.ispartofCanadian Journal of Chemistry
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectCopper selective electrodes
dc.subjectPotentiometry
dc.subjectWastewater analysis
dc.subjectChemical sensors
dc.subjectCopper
dc.subjectMetal ions
dc.subjectPotentiometers (electric measuring instruments)
dc.subjectPotentiometric sensors
dc.subjectTitration
dc.subjectVoltammetry
dc.subjectCopper-selective electrodes
dc.subjectPolyvinyl chloride membrane
dc.subjectPotentiometric response
dc.subjectPotentiometric titrations
dc.subjectTetrakis(p-chlorophenyl)borate
dc.subjectTrace-level determination
dc.subjectCopper compounds
dc.titleSolid-contact polyvinyl chloride membrane electrode based on the bis[(2-(hydroxyethylimino)phenolato]copper(II) complex for trace level determination of copper ions in wastewater
dc.typeArticle
dspace.entity.typePublication
rdl.citation.scopus22
rdl.citation.wos20
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)