Comparison of Six Methods of Central Corneal Thickness Measurement in Healthy Eyes

ArticleAçık Erişim

Özet

Objectives: Several methods of measuring central corneal thickness (CCT) have been compared, however, the data are still limited and conflicting. The aim of this study was to determine the agreement of CCT measurements performed in healthy eyes using ultrasound pachymetry (USP), non-contact tono/pachymetry, specular microscopy, biometry, Scheimpflug-based corneal topography, and optical coherence tomography (OCT).Methods: All of the participants underwent a complete ophthalmological examination. The CCT of all of the eyes included was measured using 6 different methods. The agreement between the methods was analyzed using the meandifference and Bland-Altman analysis based on a 95% limits of agreement.Results: A total of 64 patients with a mean age of 40.96±14.52 years (range: 20-78 years) were included in the study. The mean CCT value was 552.10±35.65 μm, 550.40±35.55 μm, 554.67±35.49 μm, 545.39±34.21 μm, 546.25±35.49 μm, and 552.64±33.59 μm using USP, non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflugbased corneal topography, and OCT, respectively. The bias values determined by Bland-Altman plots were -1.70, -2.56,6.71, 5.85, and -0.53 for tono/pachymetry, specular microscopy, biometry, topography, and OCT, respectively. OCT demonstrated the lowest bias compared to USP. The overall intraclass correlation coefficient was 0.961 (range: 0.945-0.974) with a 95% confidence interval.Conclusion: All of the CCT measurements obtained using non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT were consistent with the USP measurements ofhealthy controls. Larger prospective studies to determine the interchangeability of different methods for CCT measurements in pathological conditions are warranted.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/17958Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.14744/bej.2021.17894Yayıncı resmi DOI adresi
Tam Kayıt
AlanDeğerDil
dc.contributor.authorUcak, Turgay
dc.contributor.authorUĞURLU, ADEM
dc.contributor.authoriçel, erel
dc.contributor.authoryilmaz, hayati
dc.contributor.authorTAŞLI, NURDAN GAMZE
dc.contributor.authorkarakurt, yucel
dc.contributor.authorDemir, Mehmet
dc.date.accessioned2026-10-18T20:56:16.377317872Z
dc.date.available2026-10-18T20:56:16.377317872Z
dc.date.issued2021
dc.description.abstractObjectives: Several methods of measuring central corneal thickness (CCT) have been compared, however, the data are still limited and conflicting. The aim of this study was to determine the agreement of CCT measurements performed in healthy eyes using ultrasound pachymetry (USP), non-contact tono/pachymetry, specular microscopy, biometry, Scheimpflug-based corneal topography, and optical coherence tomography (OCT).Methods: All of the participants underwent a complete ophthalmological examination. The CCT of all of the eyes included was measured using 6 different methods. The agreement between the methods was analyzed using the meandifference and Bland-Altman analysis based on a 95% limits of agreement.Results: A total of 64 patients with a mean age of 40.96±14.52 years (range: 20-78 years) were included in the study. The mean CCT value was 552.10±35.65 μm, 550.40±35.55 μm, 554.67±35.49 μm, 545.39±34.21 μm, 546.25±35.49 μm, and 552.64±33.59 μm using USP, non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflugbased corneal topography, and OCT, respectively. The bias values determined by Bland-Altman plots were -1.70, -2.56,6.71, 5.85, and -0.53 for tono/pachymetry, specular microscopy, biometry, topography, and OCT, respectively. OCT demonstrated the lowest bias compared to USP. The overall intraclass correlation coefficient was 0.961 (range: 0.945-0.974) with a 95% confidence interval.Conclusion: All of the CCT measurements obtained using non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT were consistent with the USP measurements ofhealthy controls. Larger prospective studies to determine the interchangeability of different methods for CCT measurements in pathological conditions are warranted.
dc.identifier.doi10.14744/bej.2021.17894
dc.identifier.eissn2587-0394
dc.identifier.endpage13
dc.identifier.issn2459-1777
dc.identifier.issue1
dc.identifier.pmcidPMC8651033
dc.identifier.pmid35005485
dc.identifier.startpage7
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/17958
dc.identifier.volume6
dc.language.isoen
dc.relation.ispartofBEYOGLU EYE JOURNAL
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectGöz Hastalıkları
dc.titleComparison of Six Methods of Central Corneal Thickness Measurement in Healthy Eyes
dc.typeArticle
dspace.entity.typePublication
rdl.citation.trdizin0
rdl.departmentT.C. SAĞLIK BAKANLIĞI
rdl.departmentERZİNCAN BİNALİ YILDIRIM ÜNİVERSİTESİ
rdl.source.indexTRDIZIN
rdl.source.indexPUBMED