All-Solid-State PVC Membrane Isoniazid-Selective Potentiometric Sensor

ArticleAçık Erişim

Özet

In this study; a potentiometric ion-selective electrode (ISE) was developed for the determination of isoniazid. Isoniazid-Tetraphenylborate ion-pair was synthesized and the synthesized ion pair was used as ionophore substance in electrode membrane structures. For the membrane optimization, PVC membrane ion selective electrodes in different compositions were produced by using the synthesized ion pair and their potentiometric performance characteristics were investigated. It was determined that the electrode membrane with the ratio of 3% izoniazid-tetraphenylborate ion pair, 67% nitrophenyloctylether and 30% PVC performed the best potentiometric performance characteristics. For this electrode, linear range, slope, detection limit, pH working range and response time were determined as 5.0×10-6-1.0×10-2 M, 41.2 mV/decade, 2.0×10-6 M, 1-4 and ? 30 s, respectively. By using the proposed electrode, isoniazid contents of the two different drug tablets, which is used in the treatment of tuberculosis, were determined. It was seen that the results obtained by the potentiometric technique were in good agreement with those of the results obtained by UV technique at 95% confidence level

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/ebyu/18835Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorÖNDER, Ahmet
dc.contributor.authorBOZ, Hakan
dc.contributor.authorÇOLDUR, Fatih
dc.date.accessioned2026-10-18T21:03:38.950169294Z
dc.date.available2026-10-18T21:03:38.950169294Z
dc.date.issued2016
dc.description.abstractIn this study; a potentiometric ion-selective electrode (ISE) was developed for the determination of isoniazid. Isoniazid-Tetraphenylborate ion-pair was synthesized and the synthesized ion pair was used as ionophore substance in electrode membrane structures. For the membrane optimization, PVC membrane ion selective electrodes in different compositions were produced by using the synthesized ion pair and their potentiometric performance characteristics were investigated. It was determined that the electrode membrane with the ratio of 3% izoniazid-tetraphenylborate ion pair, 67% nitrophenyloctylether and 30% PVC performed the best potentiometric performance characteristics. For this electrode, linear range, slope, detection limit, pH working range and response time were determined as 5.0×10-6-1.0×10-2 M, 41.2 mV/decade, 2.0×10-6 M, 1-4 and ? 30 s, respectively. By using the proposed electrode, isoniazid contents of the two different drug tablets, which is used in the treatment of tuberculosis, were determined. It was seen that the results obtained by the potentiometric technique were in good agreement with those of the results obtained by UV technique at 95% confidence level
dc.identifier.eissn2149-4584
dc.identifier.endpage39
dc.identifier.issn1307-9085
dc.identifier.issue1
dc.identifier.startpage29
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/ebyu/18835
dc.identifier.volume9
dc.language.isotr
dc.relation.ispartofErzincan Üniversitesi Fen Bilimleri Enstitüsü Dergisi
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectKimya
dc.subjectTıbbi
dc.subjectToksikoloji
dc.subjectMühendislik
dc.titleAll-Solid-State PVC Membrane Isoniazid-Selective Potentiometric Sensor
dc.typeArticle
dspace.entity.typePublication
rdl.citation.trdizin4
rdl.departmentERZİNCAN BİNALİ YILDIRIM ÜNİVERSİTESİ
rdl.source.indexTRDIZIN